9.2 - The session time
The typical session time of a stimulation program is
30 min'. However, you can change it from 5 to 60
min' (in steps of 5), or select continuous stimulation
(the symbol C) (§ 8.2.2).
Each program offers a session time range,
according to the intended therapeutic purpose.
9.3 - Programme URGE
Pre-set parameters:
Frequency: 10 Hz. (adjustable from 1 to 20).
Pulse width: 200 µs.
Since this frequency range does not produce muscle
fatigue, there are no rest periods.
9.4 - Programme STRESS
Pre-set parameters:
Frequency: 50 Hz.
Pulse-width: 250 µs.
ACTION! time: 2 sec.
RELAX time: 4 sec.
Frequency during the rest time: 3 Hz
The 3 Hz frequency has a relaxing effect on the
pelvic floor.
9.5 - Programme MIXED
Pre-set parameters:
Frequency: 35 Hz.
Pulse-width: 200 µs.
ACTION! time: 3 sec.
RELAX time: 6 sec.
Frequency during the rest time: 5 Hz
The 5 Hz. frequency, during the RELAX time, has
the function of inhibition on the bladder muscle
(detrusor) and therefore effective for the URGE
component of the treatment.
9.6 - Programme PAIN
Pre-set parameters:
Frequency: 70 Hz.
Pulse-width: 50 µs.