The easiness for the patient to turn-ON the whole
target (the centre) will depend on the Full-Scale
(FS). The lower the FS, the easier the exercise, and
vice-versa.
Instead of the CCT target representation, it is
possible to represent the voluntary contraction
through the sidebar-graph. To change the selection
during the session, touch the area .
For this reason, if you want to select the BFB
program coming from the EVO program, you must
first select a stimulation program (󹊞, 󹊟, 󹊠, 󹊡).
The pulse-width of the stimulation programs
automatically changes from 150 to 250 µs.,
according to the frequency, and it can not be
manually adjusted.
The frequency of STRE, MIX and EVO programs
is manually adjustable from 25 to 150 Hz.
(§ 8.2.3).
By changing the frequency, ACTION and RELAX
timing will automatically adjust, by virtue of the
IntelliSTIM® system. However, you can edit them
manually (§ 8.2.4, § 8.2.5).
BEACMED s.r.l. - Via Monte Bianco, 12 - 27040 Portalbera PV (Italia)