BEACMED s.r.l. - Via Monte Bianco, 12 - 27040 Portalbera PV (Italia)
Sign
Comments /
Probable causes
Suggested remedies
12.9 - Stimulation is
no longer felt by the
patient,
All concentric circles
intermittently turn-
ON on the display
and the backlight
flashes red
(if enabled).
12.9a - The "open circuit
protection" (§ 8.1.5) intervened.
The circuit connection of the unit
to the patient is interrupted or the
contact of the electrodes with the
skin is insufficient.
12.9aa - The output circuit connection to the patient is
interrupted. Check the connection of cables and electrodes. If
required, replace the cable/s. If the wires and their connection
is ok, there may be insufficient contact between the electrodes
and the skin. Try to moisten the gel surface of the electrodes
with water. If necessary, replace the electrodes with new ones.
12.10 - No longer
feel any stimulation,
the symbol is lit
on the display and
the backlight slowly
blinks green
(if enabled)
12.10a - The device is in the state
of "PAUSE" (§ 8.2.1).
Maybe you have inadvertently
pressed the button integrated in
the upper knob.
12.10aa - Just briefly press the button to resume the
session.