EasyManua.ls Logo

BeaconMedaes MEGA3 - F = Low Alarm Settings Tab

BeaconMedaes MEGA3
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-25
MEGA3 NFPA Medical Gas Notification System
4107 9016 59.03
1. Touch the tab for “20MA=” to set the high
measurement point of the device (100
for example) and press the save button to
exit.
2. Touch the tab for “Units=” to set the
method of measurement (percent for
example), type in the measurement type
and press the save button to exit.
3. Touch the X button in the upper right to
close the screen.
F = Low Alarm Settings Tab
You can remove the low alarm set point or
customize the low alarm set point.
To remove the low alarm set point:
1. Touch the NONE tab.
2. Touch the check on the CONFIRM SET-
TINGS CHANGE screen.
To customize the low alarm set point:
1. Touch the ADJUST tab.
2. Enter the new set point.
3. Touch the Save button.
Figure 108: Units of Measure Tab
G = High Alarm Settings Tab
You can remove the high alarm set point or
customize the high alarm set point.
To remove the high alarm set point:
1. Touch the NONE tab.
2. Touch the check on the CONFIRM SET-
TINGS CHANGE screen.
To customize the high alarm set point:
1. Touch the ADJUST tab.
2. Enter the new set point.
3. Touch the Save button.
Configuring Additional 4-20mA Devices
Configure additional 4-20mA devices by
touching the number on the numeric display
that corresponds with the gas badge and
repeating the steps listed above.
Figure 109: Low Alarm Settings

Table of Contents

Related product manuals