Beckman Coulter J2-MC - User Manual

Beckman Coulter J2-MC
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
PN JR-IM-10AG
October 2014
Instructions For Use
Rotors and Tubes
For Beckman Coulter
J2, J6, and Avanti J Series Centrifuges
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea, CA 92821 U.S.A.

Table of Contents

Questions and Answers

  • W
    williamsmichaelAug 17, 2025
    What to do if my Beckman Coulter rotor lid is hard to remove?
    • J
      James MonroeAug 17, 2025
      If you're having trouble removing the rotor lid, canister cover, or bucket cover from your Beckman Coulter equipment after centrifugation, it could be due to a vacuum buildup inside the container. Try lifting the vent plug on buckets or canister covers with vents to relieve the vacuum. Alternatively, the threads may be contaminated with dirt, dried lubricant, or metal particles, or they may be insufficiently lubricated. Routinely clean metal threads with a diluted solution of Solution 555 (10 parts water to 1 part detergent), and then lubricate them with Spinkote.
  • D
    Dean DavidsonAug 20, 2025
    Why is my Beckman Coulter bottle leaking (bottles with cap assemblies)?
    • B
      brooke01Aug 20, 2025
      Leakage in bottles with cap assemblies in Beckman Coulter equipment can occur due to: * Moisture or lubrication on the cap or sealing surface. Ensure the O-ring, plug, and bottle lip are dry and free of lubrication before use. * A damaged or defective O-ring or gasket. Replace the O-ring or gasket. * Insufficiently tightened cap. Tighten the cap securely. * An uneven sealing surface on the bottle. Replace the bottle. * Overfilled tube. If using threaded caps without inserts or O-rings, keep the meniscus lower to prevent leakage.
  • D
    Dr. Laura JonesAug 22, 2025
    What to do if adapters stick in Beckman Coulter buckets?
    • V
      Victoria ShortAug 22, 2025
      To prevent adapters from sticking in buckets after centrifugation in your Beckman Coulter equipment, apply a thin film of powder, such as talcum powder, to the tube adapter rubber bases after cleaning or as required.
  • A
    Alan NobleAug 25, 2025
    How to prevent Beckman Coulter tube leakage?
    • A
      Anna HoganAug 26, 2025
      To prevent tube leakage in your Beckman Coulter equipment, ensure that caps are properly seated on tubes and fully tightened. Also, make sure all cap components are thoroughly dry before assembly.
  • H
    Heather ShepardAug 27, 2025
    What to do if Beckman Coulter Quick-Seal tubes are leaking?
    • A
      Ana GrayAug 27, 2025
      If Quick-Seal tubes are leaking in your Beckman Coulter equipment, it's likely due to improper sealing. After heat-sealing, gently squeeze the tube to test the seal for leaks. If the tube leaks, reseal it.
  • A
    Allison RobinsonAug 30, 2025
    Why is my Beckman Coulter bottle leaking (uncapped bottles)?
    • A
      Autumn VasquezAug 30, 2025
      If an uncapped bottle is leaking in your Beckman Coulter equipment, it is likely because the bottle is too full. The meniscus must be kept lower to prevent leakage.
  • R
    Richard DelgadoSep 2, 2025
    What to do if Beckman Coulter tubes with snap-on caps are leaking?
    • L
      Logan PrinceSep 2, 2025
      If tubes with snap-on caps are leaking in your Beckman Coulter equipment, ensure the tube is not too full; the meniscus must be kept lower to prevent leakage.
  • G
    Gabrielle BrennanSep 6, 2025
    Why are my Beckman Coulter tubes collapsing?
    • K
      Katrina JohnstonSep 7, 2025
      Tube collapse in Beckman Coulter equipment can occur due to several reasons: * Insufficient thinwall tube volume: The meniscus should be 2 to 3 mm below the tube top to provide adequate tube wall support. * Moisture: Ensure that tubes and tube cavities or buckets are dry before inserting the tubes, as moisture can cause the tube to float and collapse. * Reagent incompatibility: Refer to Chemical Resistances for chemical compatibilities of the tube material and chemicals used. * Excessive speed: Refer to the applicable rotor manual for maximum speeds to ensure tubes are not run above their rated speed.
  • E
    elizabeth39Sep 9, 2025
    Why are my Beckman Coulter tubes cracking?
    • J
      jeffrey11Sep 9, 2025
      Cracking in Beckman Coulter tubes can be attributed to: * Low temperature: Tubes may crack or become brittle if used below their lower temperature limit. Before using tubes at other than stated temperature limits, evaluate them under centrifugation conditions. * Aging: Tubes may become brittle with age and use. Dispose of brittle or cracked tubes.
  • D
    Douglas WrightSep 14, 2025
    What is the reason for leakage of uncapped tubes in Beckman Coulter equipment?
    • T
      thomas42Sep 14, 2025
      If uncapped tubes are leaking in your Beckman Coulter equipment, it's likely because the tube volume exceeds the maximum uncapped volume.

Summary

Safety Notice

Chemical and Biological Safety

Outlines safety precautions for handling hazardous biological and chemical materials.

Mechanical Safety

Details safety procedures related to the mechanical operation of rotors and centrifuges.

Rotors

Rotor Selection and Types

Guides users in selecting the appropriate rotor based on application and details different rotor types like fixed-angle and swinging-bucket.

General Operating Information

Covers essential operating procedures including rotor balance, tie-down, overspeed protection, and speed limits.

Tubes, Bottles, and Accessories

Filling, Sealing, and Capping

Provides instructions for filling, sealing, and capping various labware types for safe and effective centrifugation.

Sample Recovery

Explains methods for recovering samples from tubes, cups, and blood bags after centrifugation.

Using Fixed-Angle Rotors

Removal and Sample Recovery

Explains the process for safely removing rotors and recovering samples from fixed-angle rotor systems.

Using Swinging-Bucket Rotors

Rotor Preparation and Loading

Details the steps for preparing and loading swinging-bucket rotors, including balancing and special instructions for JS-24 series.

Operation and Sample Recovery

Covers operating procedures, sample recovery, and removal of JS-24 series rotors.

Using Vertical-Tube and Rack-Type Rotors

Care and Maintenance

Rotor Care Procedures

Outlines essential rotor care practices including cleaning, decontamination, inspection, and lubrication.

Beckman Coulter J2-MC Specifications

General IconGeneral
Max Speed21, 000 rpm
Temperature ControlYes
DriveDirect Drive
Temperature Range0 to 40°C
Refrigeration SystemCFC-free
Power Requirements230V, 50/60Hz