Parameter
Default
Details
Images
50
Increase value for very low concentration samples
Size
10 -40
µm
If a particular cell is being problematic, place
cursor over cell and size is displayed below
image
Brightness
85
Picks out objects from background
Increase includes more objects
Decrease includes less objects
Sharpness
100
Picks out objects from background
Decrease to include more objects
Increase to include less objects
Spot
Brightness
75 %
Determines cell viability
Increase value more cells counted as dead
Decrease value more cells counted as live
Spot Area
5 %
Ratio of the bright spot to total area
Determines cell viability
Increase value more cells counted as dead
Decrease value more cells counted as live
Minimum
Circularity
0
Excludes dead cells so they are considered debris
0 excludes no dead cells
1 excludes all dead cells
Decluster
medium
Ability of software to pick individual cells from a
cluster of multiple cells
None counts a cluster as 1 cell