5.6 Continuity Test
Prior to any continuity test, it
must be ensured that the resi-
stance to be measured is not
live. Failure to comply with
this prescription can lead to
dangerous corporal user inju-
ries or cause instrument
damage. Additionally, foreign
voltages falsify the measure-
ment result.
Position measurement function
selection switch to /
measurement range.
Connect the black test lead to
the COM socket and the red
test lead to the Temp/
/ / socket.
The resistance measurement
range is pre-set. Select the
desired measurement function
by pressing key . The
symbol is displayed on the dis-
play screen.
Connect test leads to UUT.
The measurement result is
indicated; optical display on the
screen, acoustic indication by
signal sound.
Continuity Test