The Direct Percent Entry Procedure
The direct percent entry procedure provides a way of zeroing or overwriting the electronic calibration
constant. This procedure is not accumulative. The basic and the optional calibration procedures
described previously are accumulative. The percent value entered in this procedure becomes the new
electronic calibration correction constant for the meter. This procedure requires no test can.
1. Break the seal on the electronic calibration switch and move the switch to the calibrate position.
This brings up the test can size display.
2. Press the UP-ARROW button to bring up the electronic calibration correction constant display.
3. Press the +/- button to bring up the percent error display.
4. Enter the percent error amount and press the ENTER button.
5. Move the calibration switch back to the operate position and seal it.
Viewing the Electronic Calibration Correction Constant
The electronic calibration correction constant may be viewed with the electronic calibration switch in
either position (calibrate or operate). If the switch is in the calibrate position, the correction constant can
be viewed from the test can display by pressing the UP-ARROW button. Pressing the UP-ARROW
button a second time returns to the test can display. If the switch is in the operate position, enter
manager’s mode followed by menu 27 to view the calibration constant.
How to Electronically Calibrate the SB-100 Meter (Cont’d)
(E-Cal Option)