AB 01 01/2009 page 8/26
1.4 General safety instructions on installing and operating the Alarm Box
AB 01
With regard to operating the Alarm Box AB 01and to the installation location, please observe regulations
BGR 120 issued by the German professional association for the chemical industry (formerly ZH 1/119
laboratory guidelines issued by the employers’ liability insurance association) (for Germany).
BINDER GmbH is only responsible for the safety features of the unit provided skilled electricians or
qualified personnel authorized by BINDER perform all maintenance and repair, and if components
relating to chamber safety are replaced in the event of failure with original spare parts.
CAUTION
Danger of overheating.
Damage to the unit.
∅ Do NOT install the unit in unventilated recesses.
¾ Ensure sufficient ventilation for dispersal of the heat.
The Alarm Box AB 01must NOT be operated in hazardous locations.
DANGER
Explosion hazard.
Danger of death.
∅ Do NOT operate the unit in potentially explosive areas.
¾ KEEP explosive dust or air-solvent mixtures AWAY from the unit.
DANGER
Electrical hazard.
Danger of death.
∅ The unit must NOT become wet during operation or maintenance.
The Alarm Box AB 01has been produced in accordance with VDE regulations and were routinely tested
in accordance to VDE 0411-1 (IEC 61010-1).
1.5 Intended use
The Alarm Box AB 01 has been developed for the transmission of alarm messages released by the
communication software APT-COM™ 3 DataControlSystem from BINDER, to an already existing alarm
facility of the customer, or by use of the telephone dialing device TWG 01 from BINDER via telephone
line.
Observing the instructions in this operating manual is part of the intended use.