What does it mean if my Binder Laboratory Equipment chamber is without function, and the standby icon is displayed?
- JJulie RyanSep 8, 2025
The chamber is likely in standby mode. Press down the standby button until the display lights up.
What does it mean if my Binder Laboratory Equipment chamber is without function, and the standby icon is displayed?
The chamber is likely in standby mode. Press down the standby button until the display lights up.
What to do if there is an electrical hazard during live maintenance work on Binder Accessories?
To avoid electrical hazards during maintenance on your Binder Accessories, ensure the chamber remains dry, never remove the rear panel, and always disconnect the chamber from power by turning off the main switch (for ED, FD, FED size 720 models) and pulling the plug. It is essential that only licensed electricians or experts authorized by BINDER perform any maintenance work.
How to fix a Binder Laboratory Equipment fan that doesn’t turn or turns too slowly?
If you have a BF or FED model, the fan speed might be set too low; set the fan speed to 100%. If the problem persists, further assistance is needed.
What does the alarm message “!TProt” mean on my BD or BF Binder FD 260 Laboratory Equipment?
The safety device class 3.1 has responded. Check the settings of the temperature set-point and of the safety device class 3.1.
Why are there deviations from the indicated heating-up times in my Binder Laboratory Equipment?
The chamber might be fully loaded. Try loading the chamber with fewer items or consider allowing longer heating-up times.
What to do if the Binder FD 260 Laboratory Equipment chamber has no function and the standby icon is displayed?
The chamber is likely in standby mode. Press down the standby button until the display lights up.
Requirements for qualified personnel for chamber installation, testing, and operation.
Guidelines for handling and understanding the operating manual for safe chamber use.
Explanation of safety signal words, alert symbols, and pictograms used in the manual.
Essential safety guidelines for installation and operation, including hazards.
Defines the proper use of the chambers and outlines prohibited applications.
Lists potential misuses of the chamber and associated risks.
Identifies unavoidable risks and necessary precautions during operation and maintenance.
Description of the control panel, display, buttons, and USB interface.
Detailed procedures and warnings for safely lifting and transporting the chamber.
Requirements for installation site, ambient conditions, and minimum distances.
Details on connecting the chamber to the power supply, including specifications.
Guidelines for loading the chamber, including maximum load limits and rack orientation.
Explanation of controller buttons, status symbols, and display elements.
How to set desired temperature and fan speed values for the chamber.
Explanation of the class 1 safety device that turns off heating when temperature limits are exceeded.
Details on adjustable safety controllers (class 2 and 3.1) for temperature protection.
How to set the limit temperature for the safety controller to ensure protection.
Information on alarm messages, buzzer sounds, and steps to reset alarms.
Instructions for connecting a USB stick and initiating data recording.
Steps to stop the active data recording process.
Step-by-step guide for cleaning the chamber interior and exterior surfaces.
Procedures for decontaminating the chamber from hazardous substances.
Details on recommended maintenance intervals and service procedures.
Guide to diagnosing and resolving common faults and shortcomings.
Detailed technical specifications for BD series chambers, including dimensions and electrical data.
Detailed technical specifications for BF series chambers, including dimensions and electrical data.
Detailed technical specifications for ED series chambers, including dimensions and electrical data.
Detailed technical specifications for FD series chambers, including dimensions and electrical data.
Detailed technical specifications for FED series chambers, including dimensions and electrical data.
Requirements for qualified personnel for chamber installation, testing, and operation.
Guidelines for handling and understanding the operating manual for safe chamber use.
Explanation of safety signal words, alert symbols, and pictograms used in the manual.
Essential safety guidelines for installation and operation, including hazards.
Defines the proper use of the chambers and outlines prohibited applications.
Lists potential misuses of the chamber and associated risks.
Identifies unavoidable risks and necessary precautions during operation and maintenance.
Description of the control panel, display, buttons, and USB interface.
Detailed procedures and warnings for safely lifting and transporting the chamber.
Requirements for installation site, ambient conditions, and minimum distances.
Details on connecting the chamber to the power supply, including specifications.
Guidelines for loading the chamber, including maximum load limits and rack orientation.
Explanation of controller buttons, status symbols, and display elements.
How to set desired temperature and fan speed values for the chamber.
Explanation of the class 1 safety device that turns off heating when temperature limits are exceeded.
Details on adjustable safety controllers (class 2 and 3.1) for temperature protection.
How to set the limit temperature for the safety controller to ensure protection.
Information on alarm messages, buzzer sounds, and steps to reset alarms.
Instructions for connecting a USB stick and initiating data recording.
Steps to stop the active data recording process.
Step-by-step guide for cleaning the chamber interior and exterior surfaces.
Procedures for decontaminating the chamber from hazardous substances.
Details on recommended maintenance intervals and service procedures.
Guide to diagnosing and resolving common faults and shortcomings.
Detailed technical specifications for BD series chambers, including dimensions and electrical data.
Detailed technical specifications for BF series chambers, including dimensions and electrical data.
Detailed technical specifications for ED series chambers, including dimensions and electrical data.
Detailed technical specifications for FD series chambers, including dimensions and electrical data.
Detailed technical specifications for FED series chambers, including dimensions and electrical data.
| Temperature Accuracy | ± 0.3 °C |
|---|---|
| Power Supply | 230 V, 50/60 Hz |
| Type | Drying oven |
| Temperature Range | 5 °C above ambient temperature to 300 °C |
| Interior Volume | 260 L |