Redline Operating Manual RE / RF / RI 12/2010 page 45/46
Electrical connection data redLINE UL units (for USA and Canada):
Chamber type RE 53-
UL
RE 115-
UL
RF 53-
UL
RF 115-
UL
RI 53-
UL
RI 115-
UL
Electrical data
Nominal voltage (±10%) 50/60 Hz V
100-120
V 1N~
100-120
V 1N~
100-120
V 1N~
100-120
V 1N~
100-120
V 1N~
100-120
V 1N~
Nominal power kW 0.8 1.0 0.8 1.0 0.25 0.3
Nominal current Amp 6,7 8,3 6,7 8,3 2,1 2,5
Power plug NEMA 5-15P 5-15P 5-15P 5-15P 5-15P 5-15P
Unit fuse 5 x 20 mm
250V / time-lag T
Amp 10
external
10
external
10
external
10
external
10
external
10
external
Temperature fuse class 1 (DIN 12880) internal internal internal internal internal internal
* up to 98 % of the set value
All technical data is specified for units with standard equipment at an ambient temperature of +25 °C /
77°F and a power supply voltage fluctuation of ±10. The temperature data is determined in accordance to
factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height,
width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
14.5 Equipment
To operate redLINE chambers, use only original accessories or accessories / components
from third-party suppliers authorized by the manufacturer. The user is responsible for any risk
arising from using unauthorized accessories.
Chamber type RE RF RI
Standard equipment
Microprocessor temperature controller with LED display
x x x
Integrated timer, adjustable 0-9999 minutes
x x x
Temperature safety device class 1 acc. to DIN12880
x x x
Stainless steel interior equipped with 2 chrome-plated trays
x x x
Inner glass door
-- --
x
Exhaust duct, internal diameter 50 mm / 1.97 inches, with adjustable
ventilation slide
x x x