12
CAUTION
1) Powder cap and powder seal must be attached fi rmly. An unstable attachment will cause the powder cap to undo during use
and may cause injuries or powder scattering.
2) Make sure the powder stirring cap is attached. Improper attachment will lower the cleaning effi ciency.
3) Remove powder remaining in/on the screw threads of the main body, powder cap, and sealing surface of the powder seal.
If the powder adheres fi rmly and is hard to remove, clean the remaining powder according to the instructions in “8.1 3)
Cleaning and disinfection”. As the remaining powder may cause premature wear of the threads and insuffi cient sealing
which may cause powder scattering.
4) Before each use, check if the screw threads of the main body and powder cap have any damage such as wear, crack,
deformation, etc. If the main body or powder cap has a damage, do not use and ask your dealer for repair together.
5) If the triangle symbol on the powder cap lies outside of the mounting position mark while the polisher is in use, this may
cause trouble like crack, damage or premature wear of the threads of the main body and powder cap.
ATTENTION
Remove water, oil, or powder remaining in/on the head, coupling of chamber unit, powder chamber, and powder cap.
Otherwise, the powder will get damp and cause clogging.
12
CAUTION
1) Powder cap and powder seal must be attached fi rmly. An unstable attachment will cause the powder cap to undo during use
and may cause injuries or powder scattering.
2) Make sure the powder stirring cap is attached. Improper attachment will lower the cleaning effi ciency.
3) Remove powder remaining in/on the screw threads of the main body, powder cap, and sealing surface of the powder seal.
If the powder adheres fi rmly and is hard to remove, clean the remaining powder according to the instructions in “8.1 3)
Cleaning and disinfection”. As the remaining powder may cause premature wear of the threads and insuffi cient sealing
which may cause powder scattering.
4) Before each use, check if the screw threads of the main body and powder cap have any damage such as wear, crack,
deformation, etc. If the main body or powder cap has a damage, do not use and ask your dealer for repair together.
5) If the triangle symbol on the powder cap lies outside of the mounting position mark while the polisher is in use, this may
cause trouble like crack, damage or premature wear of the threads of the main body and powder cap.
ATTENTION
Remove water, oil, or powder remaining in/on the head, coupling of chamber unit, powder chamber, and powder cap.
Otherwise, the powder will get damp and cause clogging.
13
7.5 Connecting to dental unit
1) Cleaning of coupling
(a) Confi rm no water or oil remains in the coupling on your turbine hose. Remove with a dry cloth if remains.
* When a handpiece (air turbine, air scaler, etc.) is detached, water and oil remain at the joint.
Remove the water and oil without fail.
(b) Confi rm no water or oil remains in the joint of polisher. Remove with a cotton swab if remains.
ATTENTION
Remove water or oil remaining inside of the coupling on your turbine hose and joint of the chamber unit. Otherwise, the
powder will become damp and cause clogging.
Remove water
and oil.
Remove water and oil.
13
7.5 Connecting to dental unit
1) Cleaning of coupling
(a) Confi rm no water or oil remains in the coupling on your turbine hose. Remove with a dry cloth if remains.
* When a handpiece (air turbine, air scaler, etc.) is detached, water and oil remain at the joint.
Remove the water and oil without fail.
(b) Confi rm no water or oil remains in the joint of polisher. Remove with a cotton swab if remains.
ATTENTION
Remove water or oil remaining inside of the coupling on your turbine hose and joint of the chamber unit. Otherwise, the
powder will become damp and cause clogging.
Remove water
and oil.
Remove water and oil.