8. MAINTENANCE AND CHECKS
8.1 Regular maintenance and checks
* Perform maintenance and checks as follows before sterilization and after clinic hours without fail.
ATTENTION
To use safely, check and maintain this product daily after clinic hours according to the instructions in “8. MAINTENANCE
AND CHECKS”.
1) Powder disposal
Detach the powder cap referring to “7.10 3) Detaching powder cap”. Taking care of powder scattering, discard the leftover
powder.
ATTENTION
The leftover powder must be discarded after treatment. When the powder gets damp, it will cause clogging.
2) Removal of remaining powder within polisher
(a) Attach the emptied powder cap to the chamber unit.
(b) Connect polisher to the coupling on your turbine hose.
(c) Direct the nozzle to your cuspidor. Depress your foot
control without water supply from the dental unit. Run for
1 minute to remove the powder remaining within the
channels of polisher.
(d) Disconnect polisher from your coupling.
Stop water supply.
Idle run for 1 min.
26
8. MAINTENANCE AND CHECKS
8.1 Regular maintenance and checks
* Perform maintenance and checks as follows before sterilization and after clinic hours without fail.
ATTENTION
To use safely, check and maintain this product daily after clinic hours according to the instructions in “8. MAINTENANCE
AND CHECKS”.
1) Powder disposal
Detach the powder cap referring to “7.10 3) Detaching powder cap”. Taking care of powder scattering, discard the leftover
powder.
ATTENTION
The leftover powder must be discarded after treatment. When the powder gets damp, it will cause clogging.
2) Removal of remaining powder within polisher
(a) Attach the emptied powder cap to the chamber unit.
(b) Connect polisher to the coupling on your turbine hose.
(c) Direct the nozzle to your cuspidor. Depress your foot
control without water supply from the dental unit. Run for
1 minute to remove the powder remaining within the
channels of polisher.
(d) Disconnect polisher from your coupling.
Stop water supply.
Idle run for 1 min.