Page 31 / 41 IFU_MAL_EN_V6
CAUTION
ORANGE LED TROUBLESHOOTING RECOMMENDATIONS
The Mallya base light indicator (LED) remains orange if:
- the clear square plastic window of the Mallya
base and the dose window of the pen are not
correctly aligned, or
- if the base is mounted too low / too high ( the
top of the Mallya base must be aligned with the
top of the pen, at the base of the dose button on
the pen),
- or, if the Mallya button is not pushed on enough.
In this case, it is necessary to:
- adjust and align the position of the windows:
- or to raise or lower the position of the Mallya base on the pen,
- or, to push the Mallya button onto the pen,
until the Mallya base LED flashes green and orange if Mallya is not paired (if Mallya is paired,
the light indicator is green).