Solution for precaution and warning 51
BFM-900L
FETAL MONITOR
(
Ver 1.1
)
50 Solution for precaution and warning
The beep sounds ding dong every 5 minutes when no paper until error is released or press
the Exit key. Error message will be displayed in Hi/Low and Auto Print FND until error is
released.
The beep sounds ding dong every 5 minutes when probe is disconnected during using until
error is released or press the Exit key. Error message will be displayed in Hi/Low and Auto
Print FND until error is released.
The beep sounds ding dong every 5 minutes when the battery is low until error is released or
press the Exit key. Error message will be displayed in Hi/Low and Auto Print FND until error
is released.
Display for each error
- Probe is disconnected during using : Display Pro oFF and d-1, d-2, Uc with error
code by turns in each disconnected probe.
- Probe is disconnected with no paper : Display PAP oUt and Pro oFF with error code by turns.
- Probe is disconnected with silence for 3 minutes : Display AL oFF and 3 MI with error
code by turns.
- Probe is disconnected with silence : Display AL oFF with error code.
- Probe is disconnected with paper out and silence for 3 minutes : Display AL oFF and 3
MI with error code by turns.