Solution for precaution and warning 53
BFM-900L
FETAL MONITOR
(
Ver 1.1
)
52 Solution for precaution and warning
- Probe is disconnected with paper out and silence : Display AL oFF with error code by turns.
- Paper out : Display PAP oUt and Ins Ert with error code by turns.
- Paper out with silence for 3 minutes : Display AL oFF and 3 MI with error code by turns.
- Paper out with silence : Display AL oFF with error code
- Low Battery : Display bAt Lo with error code(if internal batter is instlled : optional)
If no management with low battery within 10 minutes, the device will be turned off. Before
turning off the device, connect AC power and charge the battery. If the device is turned off
with low battery message, the device will not be turned on before connecting the AC power.