Operation
5
5000-XT
1. Make sure the sensor has reached a stable operating temperature.
2. Make sure no RF power is applied to the sensor.
3. Press and hold “0” for two seconds to begin zero calibration.
WARNING
Do not interrupt the calibration.
Note: “Zero/Cal” will be displayed and calibration will begin.
Note: Zeroing the sensor takes 60 seconds. The bar graph will dis-
play calibration progress.
Note: When complete, “PASS” should be displayed.
4. Press Enter to return to normal operation when zeroing is complete.
Note: If “FAIL” is displayed, make sure no RF power is applied to
the sensor and perform the procedure again.