Byte
offset
Meaning
3 is constant resistance (CR)
4-24 Reserved
0x3B Read the list operation (CC/CV/CW/CR)
Byte
offset
Meaning
3 List operation mode:
0 is constant current (CC)
1 is constant voltage (CV)
2 is constant power (CW)
3 is constant resistance (CR)
4-24 Reserved
0x3C Set how lists repeat (ONCE or REPEAT)
Byte
offset
Meaning
3 How lists repeat:
0 is ONCE
1 is REPEAT
4-24 Reserved
0x3D Read how lists repeat (ONCE or REPEAT)
Byte
offset
Meaning
3 How lists repeat:
0 is ONCE
1 is REPEAT
4-24 Reserved
0x3E Set the number of list steps
Byte
offset
Meaning
3 to 4 2 byte little-endian integer for number of steps
5-24 Reserved
0x3F Read the number of list steps
Byte
offset
Meaning
3 to 4 2 byte little-endian integer for number of steps
5-24 Reserved