BK Precision 8500 - Command Details - CW and CR Transient Parameter Commands; Read CW and Set CR Transient Parameter Commands; Read CR Transient Parameter and Select List Commands

BK Precision 8500
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Byte
offset
Meaning
15 Transient operation:
0 is CONTINUOUS
1 is PULSE
2 is TOGGLED
16-24 Reserved
0x38 Set CR mode transient resistance and timing
Byte
offset
Meaning
3 to 6
Value A of resistance in units of 1 . Little-endian 4 byte number.
7 to 8 Time for A resistance in units of 0.1 ms. Little-endian 2 byte number.
9 to 12
Value B of resistance in units of 1 . Little-endian 4 byte number.
13 to 14 Time for B resistance in units of 0.1 ms. Little-endian 2 byte number.
15 Transient operation:
0 is CONTINUOUS
1 is PULSE
2 is TOGGLED
16-24 Reserved
0x39 Read CR mode transient parameters
Byte
offset
Meaning
3 to 6
Value A of resistance in units of 1 . Little-endian 4 byte number.
7 to 8 Time for A resistance in units of 0.1 ms. Little-endian 2 byte number.
9 to 12
Value B of resistance in units of 1 . Little-endian 4 byte number.
13 to 14 Time for B resistance in units of 0.1 ms. Little-endian 2 byte number.
15 Transient operation:
0 is CONTINUOUS
1 is PULSE
2 is TOGGLED
16-24 Reserved
0x3A Select the list operation (CC/CV/CW/CR)
Byte
offset
Meaning
3 List operation mode:
0 is constant current (CC)
1 is constant voltage (CV)
2 is constant power (CW)
8500 DC Load Series Version: February 4, 2009 Page 59 of 76