25
Model 878A:
Resistance (parallel mode)
Test Frequency: 120Hz / 1KHz
Range Maximum
Display
Accuracy
Specified Note
@120 Hz
@1KHz
10M
9.999M
2.0%+8 *3 2.0%+8 *3 After open cal.
2000K
1999.9K
0.5%+5
0.5%+5
After open cal.
200K
199.99K
0.5%+3
0.5%+3
-
20K
19.999K
0.5%+3 0.5%+3 -
2000
1999.9
0.5%+3
0.5%+3
-
200
199.99
0.8%+5
0.8%+5
After short cal.
20
19.999
1.2%+40 1.2%+40 After short cal.
Notes:
1.
This specification is based on the measurement performed at the test socket.
2.
DUT (Device Under Test) & Test leads be properly shielded to GUARD if
necessary.
3.
This specification is based on battery operation.
Capacitance (parallel mode)
Test Frequency: 120 Hz
Range Maximum
Display
Accuracy Spec. Note
Capacitance
DF
10mF
19.99mF *5
3.0%+5
(DF<0.1)
10%+100/Cx+5
(DF<0.1)
After short cal.
1000μF 1999.9μF*6
1.0%+5
(DF<0.1)
2%+100/Cx+5
(DF<0.1)
After short cal.
200μF 199.99μF 0.7%+3
DF<0.5
0.7%+100/Cx+5
(DF<0.5)
-
20μF
19.999μF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
2000nF
1999.9nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
200nF
199.99nF
0.7%+5
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
After open cal.
20nF
19.999nF
1.0%+5
(DF<0.1)
2%+100/Cx+5
(DF<0.1)
After open cal.