Test Frequency: 10 KHz
Range
Maximum
Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
1000mH 999.9mH 2.0%+(Lx
/10000)%+ 8
2.0%+100/Lx+10 -
200mH 199.99mH 1.5%+(Lx
/10000)%+8
2.0%+100/Lx+10 -
20mH 19.999mH 1.5%+(Lx
/10000)%+10
3.0 %+100/Lx+15 -
2000
µ
H 1999.9
µ
H
2.0%+(Lx
/10000)%+10
8.0 %+100/Lx+20
After short cal.
Notes:
1.
1.Q Value is the reciprocal of DF.
2.
This specification is based on the measurement performed at the
test socket.
3.
DUT & Test leads shall be properly shielded to GUARD if
necessary.
4. Lx=counts of displayed L value, e.g. L=88.88H, then Lx=8888.