23
Inductance (Series mode)
Test Frequency: 100 / 120Hz
Range
Maximum
Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
1000H 999.9H 1.0%+(Lx
/10000) %+5
2%+100/Lx+5 After open cal.
200H 199.99H 0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5 -
20H 19.999H 0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5 -
2000m 1999.9mH 0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5 -
200mH 199.99mH 1.0%+(Lx
/10000)%+5
3%+100/Lx+5 After short cal.
20mH 19.999mH 2.0%+(Lx
/10000)%+5
10%+100/Lx+5 After short cal.
Test Frequency: 1 KHz
Range
Maximum
Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
100H 99.99H 1.0%+(Lx
/10000) %+5
2.0%+100/Lx+5 After open cal.
20H 19.999H 0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5 -
2000mH 1999.9mH 0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5 -
200mH 199.99mH 0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5 -
20mH 19.999mH 1.0%+(Lx
/10000)%+5
3.0%+100/Lx+5 After short cal.
2000
µ
H 1999.9
µ
H
2.0%+
(Lx/10000)%+5
10%+100/Lx+5 After short cal.