SAFETY INFORMATION
BIOLOGICAL SAFETY
To date, exposure to pulsed diagnostic ultrasound has not been shown to produce adverse effects. However,
ultrasound should be used only by medical professionals when clinically indicated, using the lowest exposure
times possible commensurate with clinical utility.
The ultrasonic output power of the BladderScan BVI 3000 is not user‑adjustable and is limited to the minimum
level necessary for effective performance. Data on acoustic output levels can be found in the chapter Product
Specications on page 49.
CONTRAINDICATIONS
The BladderScan BVI 3000 instrument is not intended for fetal use or for use on pregnant patients.
PRECAUTIONS & WARNINGS
Warnings indicate that injury, death, or other serious adverse reactions may result from use or misuse of the
device. Cautions indicate that use or misuse of the device may cause a problem, such as a malfunction, failure,
or damage to the product. Throughout the manual, pay attention to sections labeled Important, as these contain
reminders or summaries of the following cautions as they apply to a specic component or use situation.
To ensure safe and reliable operation for the user and patient, please heed the following warnings and cautions.
PRECAUTIONS
The BladderScan BVI 3000 and related devices may contain mineral oils, batteries, and other
environmentally hazardous materials. When the instrument and/or accessories have reached the
end of their useful service life, see the section Device Disposal on page 47.
CAUTION
To avoid paper jams, never fold the end of the paper roll or cut it diagonally or to a point.
CAUTION
Failure to adhere to the following when cleaning the control unit or the scanhead may result in
permanent equipment damage and void the instrument warranty:
Ensure that you do not immerse either component in water or a cleaning or disinfecting agent.
This may permanently damage the instrument.
Do not use Cidex
®
Plus, as it is not recommended for use with Lexan polycarbonate.
Do not subject any part of the BVI 3000 system to steam sterilization or ethylene oxide
sterilization.
CAUTION