PROCEDURE 2. ADD NOTES (OPTIONAL)
After a scanning procedure has been completed, it is possible to annotate the measurement results, and these
annotations will be included on the printout. You may access the Notes screen from the Scan Results screen.
Figure 10. Notes Screen
SEL DONE
EXAM
TYPE
ALLOWED TO FILL
POST-VOID
PATIENT CATHED
PATIENT VOIDED
NO ACTION TAKENX
Action taken
Exam type
1. If you would like to change the exam type, press the EXAM TYPE button to toggle between options
as follows:
• POST‑VOID —The measurement was taken after patient voiding. This exam type is selected by default if
the measured bladder volume is less than 100 ml.
• PRE‑VOID—The measurement was taken prior to patient voiding. This exam type is selected by default
if the measured bladder volume is between 100 ml and 400 ml.
• CAPACITY—The measurement was taken prior to patient voiding, and the bladder was lled to
maximum capacity. This exam type is selected by default if the measured bladder volume is greater than
400 ml.
SEL DONE
EXAM
TYPE
ALLOWED TO FILL
POST-VOID
PATIENT CATHED
PATIENT VOIDED
NO ACTION TAKENX
2. If you want to note the action taken as a result of the current bladder volume measurement, press the SEL
button in order to select one of the following options:
• NO ACTION TAKEN—This option is selected by default.
• PATIENT VOIDED—The patient was able to void.
• PATIENT CATHED—The patient’s bladder was emptied using a urinary catheter.
• ALLOWED TO FILL—The patient’s bladder was not as full as desired, and voiding was postponed.
SEL DONE
EXAM
TYPE
ALLOWED TO FILL
POST-VOID
PATIENT CATHED
PATIENT VOIDED
NO ACTION TAKENX