Cleaning is critical to ensuring the component is ready for disinfection. Failure to properly clean the
device could result in a contaminated instrument after completing the disinfection procedure.
WARNING
In order to maintain electrical safety, use only the provided medical‑approved power supply, battery
charger, and batteries.
WARNING
To reduce the risk of electrical shock, use only the accessories and peripherals recommended by
Verathon
®
.
WARNING
No modication of this equipment is allowed.
WARNING
The battery charger, power supply, and power cables are not intended for patient contact. Ensure
six feet (two meters) is maintained between the patient and these components.
WARNING
Risk of explosion, re, or serious injury. The BladderScan BVI 3000 system is battery‑powered.
Failure to note the following when handling the battery may result in serious injury or damage:
Never short‑circuit the battery by bringing the battery terminals into contact with any other
conductive object.
Never expose the battery to abnormal shock, vibration, or pressure.
Do not disassemble, heat above 60°C (140°F), or incinerate the battery.
Keep battery out of reach of children and in original package until ready to use.
If the battery is leaking or its case is cracked, put on protective gloves to handle it, and discard
it immediately. Always dispose of used batteries in compliance with all applicable laws and
regulations.
Put insulating tape, such as cellophane tape, on the electrodes during transportation in order to
avoid a possible short circuit, re, or electrical shock.