51
Operations & Maintenance Manual: Product Specications
Table 9. Ultrasound Acoustic Output Parameters
Values in this table are the maximum readings obtained from three test results.
ACOUSTIC OUTPUT MI
I
SPTA.3
(mW/cm
2
)
I
SPPA.3
(W/cm
2
)
Global Maximum Value 0.218* 0.0676 2.95
Associated
Acoustic
Parameter
pr.3 (MPa) 0.317
Wo (mW) 0.0676 0.238
fc (MHz) 2.12 2.12 2.12
zsp (cm) 2.80 2.80
Beam
dimensions
x
‑6
(cm) 0.367
y
‑6
(cm) 0.377
PD (µsec) 1.17 1.17
PRF (Hz) 400 400
EDS
Az. (cm) 5.46
Ele. (cm) 1.20
TIS/TIB/TIC range 0.0‑1.0*
Image Rate 5.55 Hz
M‑Lines 72
Sector Angle 120°
Sector Offset 2.035 cm
* Both MI and TI values are below 1.0.