Patient Monitor User’s Manual
2-4
unable to be connected to the AC power, the battery
can be used to supply power, and at this time it is
unnecessary to use the electrical wires, and the
instrument can be switched on directly.
The monitor can only monitor one patient at a time.
When a defibrillator is applied on a patient, the
monitor may have transient disorders in the display of
waveforms. If the electrodes are used and placed
properly, the display of the monitor will be restored
within 10s. During defibrillation, please note to
remove the electrode of limb lead to the side of the
limb.
In order to have more accurate measurements results,
the monitor should be used in quiet and comfortable
environment.
To guarantee the normal and safe operation of the
monitor, a preventive check and maintenance should
be conducted for the monitor and its parts every 6 to
12 months (including performance check and safety
check) to verify the instrument can work in a safe and
proper condition and it is safe to the medical
personnel and the patient and has met the accuracy
required by clinical use.