Patient Monitor Users Manual
9-2
Caution: If you spill liquid into the equipment of
accessories, connect you service personal or us.
9.2 Seasonal Safety Checking
Note: To ensure the performance and safety of
equipment, it must be checked after using 1 year.
When check the equipment, please contact
professional technology engineers.
Please clean the plug of power cord at least once a year. Too
much dust on plug may cause the fire.
The following safety checks should be performed at least
every 12 months by a qualified person who has adequate training,
knowledge, and practical experience to perform these tests.
The data should be recorded in an equipment log. If the
device is not functioning properly or fails any of the following
tests, the device has to be repaired.
Inspect the equipment and accessories for mechanical and
functional damage.
Inspect the safety relevant labels for legibility.
Verify that the device functions properly as described in the
)