Patient Monitor Users Manual
9-3
instructions for use.
Test the earth leakage current according IEC 60601-1 Limit:
NC 500μA, SFC: 1000μA.
Test the enclosure leakage current according to IEC 60601-1:
Limit: NC 100μA, SFC: 500μA.
Test the patient leakage current (normal operation) according
IEC 60601-1
Limit: type CF: for a.c.: 10μA, for d.c.: 10μA.
Test the patient leakage current under single fault condition
according IEC 60601-1
Limit: type CF: for a.c.: 50μA, for d.c.: 50μA.
Test the patient leakage current Mains voltage on applied
part: According IEC 60601-1:
Limit: type CF: for a.c.: 50uA.
Warning: No use-serviceable parts inside, before
servicing to authorized representative or
manufacturer.