EasyManua.ls Logo

Bosch ACS 563 - Leak Test and Pump Failures

Bosch ACS 563
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ACS 553, ACS 563 Service Manual
1 689 975 295 2021-04-13 Robert Bosch GmbH
44
Diagnostics and Testing

1. Check gauges for vacuum.
2. If pressure is over 690 mBar (10 psi), clear
hoses. If unit won’t maintain vacuum, check
the following for leakage:
Tank Liquid solenoid (S7).
Hose couplers.
Hoses.
Gauge tubing.
Suction side vacuum hoses and connec-
tions.
3. If there is no pressure on gauges and

service menu and verify low-side inlet
transducer reading. If it is reading pressure

If replacing the transducer doesn’t correct
the reading, replace the power board.
4. Refer to Vacuum Pump Does Not Run in
electrical troubleshooting.

 To ensure an accurate leak test, it
is imperative that a thorough recovery and
evacuation of the system be performed. During
the recovery process, cold spots can develop
in an automotive system. Pockets of refrigerant
in desiccant and in system oil will continue


system pressure will increase, which may be
interpreted by the unit as a leak. This will vary
somewhat with ambient temperature conditions.
1. Verify the service hoses are properly
attached to the unit (valves open) without
any restrictions.
2. If vacuum can not be maintained, using
Service Leak Check in the service menu,
pressurize the unit and use a leak detector

for leaks. Repair as needed.
3. Check the oil inject (S14), tank Liquid (S7)
and vacuum (S9) solenoids for bleed-
through. Repair as needed.
 A leaking tank liquid solenoid (S7) may

zero. A leaking vacuum (S9) or oil inject (S14)
solenoid would not normally rise above zero.
4. Verify the low-side pressure transducer is
functioning properly. Enter service menu
and verify low-side inlet transducer reading.
Compare the low-side transducer reading
shown in the service menu with the
reading on the low-side panel gauge. The
readings should match, assuming the
low-side panel gauge is accurate.
If the low-side inlet transducer is reading
pressure while the system is in a vacuum,
replace the transducer. Verify the
replacement transducer reads zero at
atmosphere by plugging it into the wiring
harness prior to installation.
If replacing the transducer doesn’t correct
the reading, replace control board.

Table of Contents

Related product manuals