26 | BAT 645 | Faults & Troubleshootingen
5.6 Measuring points on the power module
DANGER - Risk of electric shock from live parts!
Work on electrical systems or equipment must only be performed by qualified electricians or
trained personnel under the guidance and supervision of an electrician.
Observe the electrical engineering safety regulations.
Voltage measurement From To Value, approx.
Measure the voltage after the power inlet socket connector X4 X2 230V
Measure the voltage after the device fuse M1 M2 230V
Measure the link voltage (UZK) C58+ C58- 360V
Measure the 1st primary auxiliary voltage VDD GND 13V
Measure the 2nd primary auxiliary voltage VDD2 GND 12V
Measure the auxiliary voltage IC 14 Pin 1 GND1 Pin 4 12V
Measure the controller voltage 3V3 GND1 Pin 2 3.3V
Measure the reference voltage 3V3ref GNDA 3.3V
Voltage on the power module with a 12-V battery U
ist
= 12V GNDA 0.85V
Voltage on the power module with a 24-V battery U
ist
= 24V GNDA 1.7V