9.2 Setting the calibration interval
9.3 Setting the calibration gas com-
ponents
9.4 Setting the calibration sequenz
In this menu you can enter the interval at
which the BEA must be recalibrated with
calibrating gas.
! Observe legal requirements when
entering the calibration interval!
Once you have made these entries,
please refer to Sections 6.4 and 6.5.
Enter the calibration interval using the
keyboard, e.g. 365.
Confirm your entry with E or by clicking
OK.
The OK box is displayed for 1 s.
The calibration interval setting function is
now complete.
This menu allows you to define which gas
measuring channels must be recalibrated
with calibrating gas during the yearly
recalibration process.
! Observe legal requirements when
making your entry!
i Enter the gas measuring channels in
capital letters.
The components you have
preselected must be contained in the
calibrating gas used for recalibration.
Using the keyboard, enter which gas
measuring channels are to be
recalibrated.
Confirm your entry with E or by clicking
OK.
The OK box is displayed for 1 s.
Entry is now complete.
This menu enables you to define whether,
once the calibration interval has elapsed,
the BEA displays a message permitting
further work or whether a message is
displayed and the BEA is subsequently
disabled and no further work permitted.
! Observe legal requirements when
making your entry!
Select the appropriate setting with the o
and u keys and confirm with E or by
clicking Next.
The OK box is displayed for 1 s.
Entry is now complete.