EasyManua.ls Logo

Bosch CISS - Page 9

Bosch CISS
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
4.11.4 Environmental sensors - temperature, pressure, relative humidity – BME280
For the environmental measurands - temperature, relative humidity and air pressure - a BME280 is used. With the influence
of the housing and self-heating of the electronic components, the behavior given in the table below has been characterized
for the CISS. The given accuracy is specified with the following measurement setup:
One environmental measurement per minute
No accelerometer, magnetometer, gyroscope active
Power supply via the CISS USB cable
TEMPERATURE
Range -20 – 80 °C
Resolution 0.1 °C
Precision ± 0.1 °C
Response time (τ
63 %
) from 20 °C to 50 °C
at 40 % r.H. and 980 hPa (with medium
access)
7.5 minutes
Accuracy max. ± 2 °C +3% T °C
Sampling rate BLE / UBS: ≤ 1 Hz
PRESSURE
Range 300 – 1100 hPa
Resolution 0.01 hPa
Precision (1-sigma level) ± 0.01 hPa
Response time < 1 s
Accuracy at 25 °C and 40 % r.H. ± 1.5 hPa
Sampling rate BLE / UBS: ≤ 1 Hz
HUMIDITY
Range 0 – 90 % r.H.
Resolution 0.01 % r.H.
Precision (1-sigma level) ± 0.5 % r.H.
Response time (τ
63 %
) from 80 % r.H. to 40
% r.H. at 25 °C and 980 hPa
20 minutes
Accuracy at 20 – 80 % r.H. max. ± 7% at +20 °C
max. ±10% at -20 °C
Sampling rate BLE / UBS: ≤ 1 Hz