EasyManuals Logo

Boston Scientific Vercise DBS Series User Manual

Boston Scientific Vercise DBS Series
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
Stimulator Implantation
Physician Manual
91098825-02 29 of 337
en
7. Grasp the Tip rmly with one hand and, while holding the Straw in place with the
other hand, pull the Shaft out of the Straw.
8. Push the proximal ends of the DBS Extensions through the Straw, and then
withdraw the Straw.
9. Optionally secure the DBS Extension connector to the fascia using sutures and/or
suture sleeves.
CAUTION: Do not use polypropylene sutures as they may damage the suture
sleeve. Do not suture directly onto the DBS Extension or use a
hemostat on the DBS Extension body. This may damage the DBS
Extension insulation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Boston Scientific Vercise DBS Series and is the answer not in the manual?

Boston Scientific Vercise DBS Series Specifications

General IconGeneral
BrandBoston Scientific
ModelVercise DBS Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals