EasyManua.ls Logo

Boston Scientific Vercise DBS Series

Boston Scientific Vercise DBS Series
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Physician Manual
Physician Manual
91098825-02 52 of 337
en
2. Press P or to move until the program you want is highlighted.
3. Press ▲ to activate the program.
Once you have selected and activated a program, the Remote Control will return to
the Level screen.
Modifying and Saving Programs
1. Press P as needed from the Level screen to cycle to the desired program.
Note: The most recently retrieved or saved program will be underlined(_).
Empty program slots are denoted by a blank box with no program number [].
If a patient tries to activate an empty program, nothing will happen.
2. Press ▲ to activate the selected program.
3. After the Remote Control times out to the Level screen, you may change the
amplitude of all areas in the program.
4. To save/store changes, select the program again and press ▼. You will be
required to conrm the adjustment rst.
5. Press P to conrm, or press the button to cancel the operation.
Note: To save the modications as a new program, simply select an empty []
program slot and press ▼ instead of overwriting the existing program.

Table of Contents