Physician Manual
Physician Manual
91098825-02 48 of 337
en
The Remote Control will display the following message upon rst activation (when
any button is pressed):
1. Ensure the Remote Control is within telemetry range of the Stimulator (60 cm
or 2 ft).
2. Press P to initiate communication between the Remote Control and the
Stimulator. The Remote Control will identify the Stimulator by ID number.
3. Press P to conrm and continue.
The Remote Control will automatically detect available program sets in the Remote
Control or Stimulator. If there are no available programs, the link will be completed
immediately and you will see the screen below:
However, if programs are present in either device (Remote Control or Stimulator)
during linking, you will be required to respond to one or more decision screens.
Your decisions will guide the Remote Control to complete the linking with the
desired program set saved to the desired device (Remote Control or Stimulator).
You may also need to enter the clinician’s password.
If an error occurs during the process or if the password is incorrect linking will
be aborted. For further information, including information about the clinician’s
password, see Selection #3 – Clear Link in the Clinician Options section.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Boston Scientific Vercise DBS Series and is the answer not in the manual?

Boston Scientific Vercise DBS Series Specifications

General IconGeneral
BrandBoston Scientific
ModelVercise DBS Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals