- 22 -
Using Control Solution
H
3,5mL
3
olllösung
ol solution
on de contrôle
plossing
one di controllo
test
®
3 Control H
The Omnitest® 3 Control solution is used to check that the
blood glucose monitoring system consisting of the meter and the
test strips is working correctly.
The control solution should be used:
Whenever you suspect the meter or test strip is not
functioning properly.
If your blood glucose test results are not consistent
with your symptoms or if you think they are not accurate.
If you have dropped the meter.
For quality control in the point of care usage.
For teaching or learning the system.
To perform a control solution test, follow the same test pro-
cedure as for a blood glucose test, substituting control solution
for the drop of blood. For detail instruction, see page 46
“Testing Your Blood Glucose”.
!
NOTE
Use only Omnitest® 3 Control
solution for the Omnitest® 3
monitoring system. The Omnitest® 3
Control solution is sold separately.