EasyManua.ls Logo

BRIGHT OTF 5000 - Quick-Release Type Holders; Orientating Specimen Holders

BRIGHT OTF 5000
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OTF 5000/HS/LS-001/2/3/4 Iss 4 June 2012
29
there is not an accumulation of rime present on the underside and leading edge of the
anti-roll plate. Care must be taken when using Cryospray 134. The risk of creating an
aerosol of potentially infectious tissue particles must be borne in mind.
5 If the section is skewing, this is a sign that the anti-roll plate is not parallel to the edge of
the knife. Remember also that the angle of the underside of the anti-roll plate must be
parallel to the facet of the knife (but just a little larger for the best results). If the angle is
too large then the section will curl. If the angle is too small then the section will crumple.
6 If there is any doubt that the temperature of the microtome or cryostat chamber is not at
the right temperature, be patient and wait for a while until the temperatures equalise.
7 For other problems follow the advice in section 3.6 below.
3.6 FROZEN SECTIONING - A PROBLEM SOLVING GUIDE
Cutting good frozen sections requires skill and practice, and some tissues are particularly
difficult. In all cases the quality of the result depends on a number of factors.
The notes which follow deal with the more commonly encountered difficulties, and suggest a
systematic approach to problem solving.
3.6.1 SOLUTIONS - CHECK LIST
First check all the basic factors:
Specimen
was it frozen badly?
has it equilibrated to chamber temperature?
has it become loose on the specimen holder?
is the specimen holder tightly clamped?
is the orientation lever locked?
has the specimen dehydrated?
Knife
has it become dull?
is it nicked or chipped?
is the knife tilt angle incorrect?
is the knife loose?
is there a build-up of frost or debris on the knife?
Anti-Roll Plate
is it set too high/low?
is the angle to the knife too wide/narrow?
is it parallel to the knife?
is it frosted/too warm?
is it damaged? (See section 2.2.8).
General Factors
is the chamber temperature appropriate to the specimen (or specimen temperature
control, where fitted)?
is the cutting motion too fast/slow/uneven?
has the microtome reached the end of its travel?
is the section thickness setting appropriate?
3.6.2 SPECIFIC PROBLEMS
Thick/thin sections, intermittent failure to cut:
knife not sharp enough
knife tilt angle too high/low
clamping screws too loose
specimen loose on holder
cutting temperatures too warm/cold
anti-roll plate adjusted too high
tissue expansion due to block warming up
Sections crumbling or not forming:

Table of Contents

Related product manuals