AMETEK Brookeld Page 13 Manual No. M15-356-B0916
In the Options screen, the RANGE key selects the currently displayed option or viscosity unit
and/or exits the Options screen.
SPEED +
Used to scroll UP (in an increasing value direction) through the available speed tables.
SPEED -
Used to scroll DOWN (in a decreasing value direction) through the available speed tables.
SPINDLE (UP)
Used to scroll UP (in an increasing value direction) through the available spindle tables. In
the Options screen, used to scroll UP through available options and viscosity units.
SPINDLE (DOWN)
Used to scroll DOWN (in a decreasing value direction) through the available spindle tables.
In the Options screen, used to scroll DOWN through available options and viscosity units.
I.8 PreventaveMaintenanceandCleaning
Make sure the instrument is in a clean, dry working environment (dust-free, moderate
temperature, low humidity, etc.).
Make sure the instrument is on a level surface.
Hands/ngers must be clean and free of residual sample. Not doing so may result in deposit
build up on the upper part of the sha and cause interference between the sha and the
pivot cup.
Be sure to remove spindle from the instrument prior to cleaning. Note le-hand thread.
Severe instrument damage may result if spindle is cleaned in place.
When cleaning spindles, do not apply excessive force, which may result in bending the
spindle sha.
Cleaning Instrument, Keypad and Display:
Clean with dry, non-abrasive cloth. Do not use solvents or cleaners.
Immersed Components:
Spindles and guard leg are made of stainless steel. Clean with non-abrasive cloth and solvent
appropriate for sample material that is not aggressive to stainless steel.