Reference Values
18 / 96 DOC-M88-EXX157_V17_04.2019
Table3.4: Maximum allowed values for the line shape in terms of FWHM (Co-radiation). SC: scintillation
counter, SOL‑XE: energy dispersive solid state detector, position sensitive detector VÅNTEC‑1,
SSD160, LYNXEYE/ LYNXEYEXE/ LYNXEYEXE‑T, LYNXEYE-2, SSD160-2 and EIGER2R500K
Reflection
hkl
Bragg-Brentano Parallel Beam
(Göbel Mirror)
SC/SOL‑XE
[°]
VÅNTEC‑1
[°]
LYNXEYE/
LYNXEYEXE/
LYNXEYEXE‑T/
SSD160/
LYNXEYE‑2/
SSD160-2
EIGER2R500K
[°]
SC/ SOL‑XE/
LYNXEYE/
LYNXEYEXE/
LYNXEYEXE‑T/
SSD160/
LYNXEYE‑2/
SSD160-2
1)
EIGER2R500K
2)
[°]
0 1 2 0.06 0.10 0.06 0.22
1 0 4 0.06 0.10 0.06 0.22
3 0 0 0.10 0.11 0.10 0.22
2 2 6 0.18 0.18 0.18 0.27
1)
LYNXEYE / LYNXEYE XE / LYNXEYE XE-T / SSD160/ LYNXEYE‑2/ SSD160‑2 in 0D
mode
2)
EIGER2R500K in 0D mode with 14mm*16mm opening
For Parallel Beam Geometry the 0.2° equatorial Soller is used as standard and optimum set
up. If different equatorial Sollers are used, FWHM values will increase accordingly.

Table of Contents

Other manuals for Bruker D8 Discover

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker D8 Discover and is the answer not in the manual?

Bruker D8 Discover Specifications

General IconGeneral
BrandBruker
ModelD8 Discover
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals