7
Z33048_00_01
Getting Started
1.6 Adjusting the Air Pressure
Adjust the air pressure so that the manometer in the console reads between 3.5 and 4
bar. This way the default spin and lift parameters will be sufficient to get you started.
1.7 Switching On the Console the First Time
The console has three modes:
• OFF (blue light OFF), the switch on the rear panel is set to OFF.
• STAND BY (blue light BLINKING), the switch on the rear panel set to OFF, the ON/
OFF button on the top of console is not yet pressed.
• ON (blue light ON).
It is IMPORTANT to switch the console on BEFORE switching on the PC, since the con-
sole contains the Ethernet hub, and the DHCP server will not start if a network is not
found.
1.7.1 Verifying That the Console Has Booted
This procedure takes 30-60 seconds, provided the DCHP server is already running.
There is no visual indication on the console indicating that the console has booted, but
the easiest way to check is to see if the three main Ethernet devices (main board, lock
and VTU) can be reached via ping or a Web browser. Their addresses will usually be in
the range:
149.236.99.25x
To access any of the Web server sites, the following password and user name are
required:
User: root
Password: BRUKER (in capitals!)
If you are not sure the DHCP Server is running, you can check by opening:
Control Panel -> Administrative Tools -> Services -> Bruker DCHP Server
Figure 1.2 Bruker DCHP Server