Loading a sample
Check your solvent height. For best shimming homogeneity, the sample height should be 5 cm.
Wipe the exterior of the tube clean using a Kimwipe.
If present, remove the cap from the top of the magnet bore.
Type “ej” in the command line at the bottom of the screen to turn the lift on.
Use the depth gauge to adjust the spinner height on the tube.
Carefully place your sample into the bore. Make sure air flow is on before releasing sample!
Type “ij” to turn off the lift.