Introduction
10 / 48 H171764_6_001
2.8 Operating Environment
Permissible Ambient
Temperature:
5 to 35 °C.
Permissible Altitude: Up to 2000 meters above sea level.
Relative Humidity: A maximum of 80% up to 31 °C and linearly
decreasing to 65% at 35 °C.
Permissible Storage Temperature: 5 to 40 °C
Ingress Protection Class: IP 20
Table2.1: Spectrometer System Operating Environment
The power requirements for the various spectrometer systems vary according to the
configuration. Additional information on power requirements can be found in the
corresponding Site Planning Manual.
2.9 Signs and Labels
Signs and labels always relate to their immediate vicinity. The following signs and labels are
found on or around the spectrometer system:
Prohibition sign: No person with pacemakers!
• People with pacemakers are endangered in the marked area and
are not allowed to enter these areas!
Prohibition sign: No person with implants!
• People with metallic implants are endangered in the marked area
and are not allowed to enter these areas!
Prohibition sign: No pregnant person.
• Pregnant people may be endangered in the marked area and
should not be allowed to enter these areas!
Prohibition sign: No watches or electronic devices!
• Watches and electronic devices may be damaged in the marked
area!
Prohibition sign: No credit cards or other magnetic memory!
• Credit cards and magnetic memory may be damaged in the marked
area!.