Cabinet Safety
20 / 48 H171764_6_001
3.6.1.1 Safety Labels
Labels are provided on amplifiers to alert operating and service personnel to conditions that
may cause personal injury or damage to the equipment from misuse or abuse. Users should
read the labels and understand their meaning.
Operating personnel should not remove RF output cables without checking whether an
experiment is running. To be sure that no RF signal is generated, type stop on the TopSpin
command line, or click on the STOP icon in the TopSpin menu bar. When in doubt, turn off
the RF transmitter.
Disconnect the mains power cable before opening the unit to prevent risk of electrical shock.
3.6.2 Emergency Shutoff
In case of an emergency, use the mains switch described in the console Emergency Shutoff
[}î16] to switch off the complete console. The amplifier in AVANCE NEO systems can only be
switched off through the PDU. It is not advised to selectively switch off the amplifier. Instead,
shutdown the console completely. Only if otherwise instructed by Bruker personnel, you may
be guided through a process to selectively shutdown the amplifier.