BTL-2000
USER'S MANUAL & USER’S GUIDE
Page 28 z 45
3 TECHNICAL PARAMETERS
3.1 Technical parameters BTL-2000
General Parameters
- Weight 700g ±50g
- Dimension 234 x 132 x 33 mm ±3%
- Cover (by EN 60529) IP20
- Class (by EN 60601-1) II (with inner power source and additional external source)
- Class (by 93/42EHS) Iib
- Type of applied part BF
- Display LCD display with backlight
- Modes of operations Continual
Operating conditions:
- Temperature 10 °C to + 40 °C
- Relative humidity 30 % to 75 %
- Atmospheric pressure 700 hPa to 1060 hPa
- Position N/A
Storage conditions
- Temperature - 10 °C to + 55 °C
- Relative humidity 25 % to 85 %
- Atmospheric pressure 650 hPa to 1100 hPa
- Position N/A
- Maximum Duration 1 year
Transport conditions
- Temperature - 10 °C to + 55 °C
- Relative humidity 25 % to 85 %
- Atmospheric pressure 650 hPa to 1100 hPa
- Position N/A
- Maximum Duration 1 year
Power supply
- external Adapter BTL-282 (see 3.2 Technical parameters ADAPTER
BTL-282)
Battery Supply
- battery type NiCd accumulator
- number 8
- total voltage 9.6 V
- capacity 950 mA
- min. operational time 1.5 h (probe 300 mW + backlight)
Voltage Insulation
- Mains Power to Patient Circuit
Insulation
1.5 kV
- Frequency * *** 0.0 – 10.0 Hz, step 0.1
11.0 – 100.0 Hz, step 1
110 – 1000 Hz, step 10
1100 – 5000 Hz, step 100
- Standard Deviation ± 3 % of default value
- Density * 0.1 – 10.0 J/cm2, step 0.1
11.0 – 99.0 J/cm2, step 1