EasyManua.ls Logo

Buchi Pure C-810 - Finding Separation Conditions with Navigator

Buchi Pure C-810
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Büchi Labortechnik AG
Appendix | 10
Operation Manual Pure Chromatography Instruments
97/102
10.3 Finding out separation conditions with the navigator
10.3.1 Open the navigator
Navigation path
Gradient panel
u Navigate to the
Gradient
panel according the navigation path.
u Tap the button [Navigator].
ð The display shows the dialog
Navigator
.
10.3.2 TLC Silica
The TLC-Silica tool uses TLC separation data to provide a recommended gradient for
silica separations of 2 or 3 components.
Specification of the samples needed:
Two different solvent concentrations
Two separations
Silica TLC plates
Precondition:
R The display shows the dialog
Navigator
. See Chapter10.3.1 "Open the
navigator", page97
u Select the
TLC- Silica
tab.
u Select the radio button for two or three components.
u Select the weaker solvent at drop-down list A.
u Select the stronger solvent at drop-down list B.
u Enter the solvent B concentrations used in the TLC separations.
All manuals and user guides at all-guides.com

Table of Contents

Related product manuals