4 Description of function
27 R-210/215 Operation Manual, Version B
4.1.3 Rear connections of the Rotavapor
1
2
3
4
Power supply for Rotavapor
Socket for control cable to the vacuum
controller (RS-485) or the Vacuum Pump V-
700/710 (to be used alternatively)
Socket for vapor temperature sensor
Power supply for heating bath
Fig. 4.3: Rear connections of the Rotavapor
4.2 Motorized quick-action jack
1
With the motorized quick-action jack, the eva-
porating fl ask can be lowered into, and raised
from, the water bath. This operation requires no
physical effort. Motion is effected by an integral
motor. The button which can be pushed
up or down for the corresponding direction is
located on the handle of the quick-action jack.
When the instrument is switched off manually or
if a power failure occurs, the fl ask is automati-
cally raised from the heating bath.
Fig. 4.4: Quick-action jack