4 Description of function
31 R-210/215 Operation Manual, Version B
4.10 Infrared interface
1
2
Receiver
Transmitter
Fig. 4.12: Infrared interface
The infrared interface transmits heating bath data to the controller. An interruption of the communica-
tion with the interface does not affect the process.
Function of the infrared interface:
Transmits the current heating bath temperature to enable working with the solvent table.
Reads out the heating bath data (set / actual temperature) and transmits them to the Rotavapor.
From there, the data is transmitted via the RS-485-port to the vacuum controller (V-850/V-855).
•
•