EasyManua.ls Logo

Buckleys DCCT PRO2 - Risk Assessment; Factory Settings

Default Icon
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual
6
06/23
Risk Assessment
It is the user’s responsibility to complete a risk assessment before using Buckleys equipment. The following
points offer some guidance but must not be assumed to be complete or sufficient.
Personal Safet
y
Have all users been trained in the correct
and safe use of the instruments?
Are they aware that the instrument
produces high voltages, and can give
electric shocks if used incorrectly?
Is the instrument in good condition,
undamaged? Is the instrument dry? Are the
accessories similarly in good condition, dry
and undamaged?
Is it possible to ensure that persons with
pacemakers, cochlear implants or trans-
dermal implants of any kind are sufficiently
far from the instrument & test site to be safe?
Ozone is an irritant gas which will be
produced when the instrument produces
sparks – is there sufficient, suitable
ventilation in place to ensure that this is not
hazardous?
Is the access route to the area of the test
safe? Is fall protection necessary? If so, is it
in place?
Has the instrument been properly grounded
to a metallic grounding point?
Should you use the anti-static wristband to
prevent static charge build-up?
Explosion & Fire Risk
Buckleys high-voltage instruments will
produce sparks - Are you certain that there
is no r
isk of an explosive atmosphere
?
Have all easily-ignited and/or flammable
mater
ials been removed from the ar
ea in
which
test
ing will take place?
EMC & RFI
Buckleys high-voltage instruments will
pr
oduce sparks which may create
interference in nearby electronic appara
tus,
partic
ularly communication and comp
uter
equipment.
Are you certain that there is no risk of such
inter
ference causing a danger to other
s –
partic
ularly medical or li
fe support
equipment?
Are you sure that there is no risk of such
inter
ference causing costs or inconvenien
ce
to others – e.g., industrial processes,
me
asuring equipment, sc
ientific apparatus,
domes
t
ic radio and TV?
Other Risks
Have you checked that there are no
ongoing pr
ocesses in the ar
ea that may be
hazar
dous?
Are you certain that the instrument has not
b
een immersed, dropped or otherwi
se
damaged i
n a way that is not obvi
ous?
Have you verified that the ground
connection
is actually grounded
?
Is the person responsible for site safety
aware of / f
ully understanding the test
ing
you ar
e going to do, and has it be
en
r
eviewed in line with the site procedur
es?
Factory Settings
In order to use the DCCT Pro2, you will need to adjust the instrument settings to suit your specific
application – see page 9. The following is a summary of the settings as set in the factory:
Output voltage: 5kV
Testing standard: Manual
Sensitiv
ity: 200µA
Al
arm volume: 5
Screen contrast: 5
Alarm Mode: Fast
ON time 30ms
OFF time 30ms
HV Pause 250ms
This setting is suitable for tests carried out in line with the BS standard
Electric Cables – Spark-test
method
.
NOTE: Wherever you are intending to use
equipment of this type, always obtain
clearance from the company safety officer.

Related product manuals