Useful Advice
Instrument sensitivity
Several variables including the test material itself, temperature, solar load and electrode type and size
will impact the Alarm Threshold setting required.
As noted in the Basic Physics section above, the detector will show a low-level current draw in micro-
amps (µA) as the electrode is passed over a coating. A threshold setting around 100µA above the
displayed current draw is often suitable; although attention should be given to verify the effectiveness
of the sensitivity setting; too low may cause false-alarms and too high could allow faults to go
undetected.
It may not always be possible to determine a suitable sensitivity setting. In such cases, increasing the
sensitivity until the detector switches to ‘Spark Only’ mode prevents the alarm from triggering unless a
direct connection to ground is made.
A higher test voltage may be required to compensate for current drawn as the electrode passes over
the coating. This method is less precise and will not detect weak or thin areas as readily. It is
recommended to consult with Buckleys in this situation as a multi-channel approach may provide an
optimum solution to the issues of excess quiescent current draw.
This generation of instrument is capable of detecting flaws much more rapidly than previous models.
Product registration
Thank you for choosing a Buckleys product,
we are sure it will provide you with many years
of reliable service.
Please register this product via Buckleys’
website and download the
Warranty Registration Certificate.
Register your product in 5 minutes
Once your product is registered, you will receive
the following benefits:
• FREE annual service & calibration reminders by email
• Latest industry news relating to your product
• Be the first to hear about our new products
We strive to improve the quality of our products and service.
Registering your product helps us monitor overall quality of our products, service and dealer network.
Additionally, if we ever need to contact you regarding your product, we are able to do so immediately.
We will also send you annual service/calibration reminders by email to help ensure your product is always in
perfect working order.
To register your product, simply scan the QR code or visit:
www.buckleysinternational.com/services/registration
...Complete the online form and click on SUBMIT.